In our study, the Icare ic200 overestimated the IOP. The overestimation increased as the baseline IOP increased. The agreement between the IOP measurement by GAT and Icare ic200 was <2 mm Hg at all ranges of IOP. The narrow limits of agreement between the tonometers for an IOP <21 mm Hg makes it a useful alternative to GAT in this pressure range.
-
There was excellent agreement between the IC200 and GAT tonometers, both in healthy subjects and PCG, with a trend to overestimate IOP when measured with IC200. There was no influence by CCT on IOP measurements in patients with PGC.